diff options
Diffstat (limited to '')
-rw-r--r-- | Documentation/devicetree/bindings/iio/health/afe4404.txt | 29 |
1 files changed, 29 insertions, 0 deletions
diff --git a/Documentation/devicetree/bindings/iio/health/afe4404.txt b/Documentation/devicetree/bindings/iio/health/afe4404.txt new file mode 100644 index 000000000..0b52830a0 --- /dev/null +++ b/Documentation/devicetree/bindings/iio/health/afe4404.txt @@ -0,0 +1,29 @@ +Texas Instruments AFE4404 Heart rate and Pulse Oximeter + +Required properties: + - compatible : Should be "ti,afe4404". + - reg : I2C address of the device. + - tx-supply : Regulator supply to transmitting LEDs. + - interrupts : The interrupt line the device ADC_RDY pin is + connected to. For details refer to, + ../interrupt-controller/interrupts.txt. + +Optional properties: + - reset-gpios : GPIO used to reset the device. + For details refer to, ../gpio/gpio.txt. + +Example: + +&i2c2 { + heart_mon@58 { + compatible = "ti,afe4404"; + reg = <0x58>; + + tx-supply = <&vbat>; + + interrupt-parent = <&gpio1>; + interrupts = <28 IRQ_TYPE_EDGE_RISING>; + + reset-gpios = <&gpio1 16 GPIO_ACTIVE_LOW>; + }; +}; |